dc.description.abstract |
Conclusions: TTD values were longer for the DL system than for the two commonly used blood culture systems when tested on simulated blood cultures. Thus, clinical laboratories considering the DL system should take its long TTD into consideration. |
|
dc.language |
English |
|
dc.publisher |
CLIN LAB PUBL |
|
dc.subject |
Medical Laboratory Technology |
|
dc.title |
Time-to-Detection Comparison for a Novel Blood Culture System Using Simulated Blood Cultures: DL (TM) versus BacT/ALERT (TM) and BACTEC (TM) |
|
dc.type |
Article |
|
dc.identifier.volume |
62 |
|
dc.identifier.startpage |
1595 |
|
dc.identifier.endpage |
1600 |
|
dc.contributor.department |
Sakarya Üniversitesi/Tıp Fakültesi/Temel Tıp Bilimleri Bölümü |
|
dc.contributor.saüauthor |
Köroğlu, Mehmet |
|
dc.contributor.saüauthor |
Altındiş, Mustafa |
|
dc.relation.journal |
CLINICAL LABORATORY |
|
dc.identifier.wos |
WOS:000383317300001 |
|
dc.identifier.doi |
10.7754/Clin.Lab.2016.160110 |
|
dc.contributor.author |
Tayfur Demiray |
|
dc.contributor.author |
Köroğlu, Mehmet |
|
dc.contributor.author |
Altındiş, Mustafa |
|