Açık Akademik Arşiv Sistemi

FAAS Determination of Au and Pd After Selective Preconcentration on Ethylenediamine Silica Material Prepared By the Sol-Gel Method

Show simple item record

dc.contributor.authors Imamoglu, M; Albayrak, S;
dc.date.accessioned 2020-02-24T14:15:16Z
dc.date.available 2020-02-24T14:15:16Z
dc.date.issued 2012
dc.identifier.citation Imamoglu, M; Albayrak, S; (2012). FAAS Determination of Au and Pd After Selective Preconcentration on Ethylenediamine Silica Material Prepared By the Sol-Gel Method. ATOMIC SPECTROSCOPY, 33, 35-30
dc.identifier.issn 0195-5373
dc.identifier.uri https://hdl.handle.net/20.500.12619/44862
dc.description.abstract The recoveries for Au(111) and Pd(II) ions under optimum preconcentration conditions were 99 +/- 2 and 98 +/- 2% at the 95% confidence level, respectively. The analytical detection limits of Au and Pd were 0.20 and 0.21 mu g L-1, respectively. The presented method was validated by the analysis of certified reference platinum ore (SARM7B). The obtained results agreed with the certified values.
dc.language English
dc.publisher PERKIN-ELMER CORP
dc.subject Spectroscopy
dc.title FAAS Determination of Au and Pd After Selective Preconcentration on Ethylenediamine Silica Material Prepared By the Sol-Gel Method
dc.type Article
dc.identifier.volume 33
dc.identifier.startpage 30
dc.identifier.endpage 35
dc.contributor.department Sakarya Üniversitesi/Fen-Edebiyat Fakültesi/Kimya Bölümü
dc.contributor.saüauthor İmamoğlu, Mustafa
dc.relation.journal ATOMIC SPECTROSCOPY
dc.identifier.wos WOS:000301404900005
dc.contributor.author İmamoğlu, Mustafa


Files in this item

Files Size Format View

There are no files associated with this item.